Pribolab - STD#7072UPribolab®交链孢霉属毒素混合内标(U-[13C14]-AOH,U-[13C15]-AME,U-[13C22]-TEN,U-[13C10]-TEA)/甲醇 (AOH 6ppm、AME 0.6ppm、TEN 3µg/mL,TEA 15 µg/mL-42347

Pribolab®交链孢霉属毒素混合内标(U-[13C14]-AOH,U-[13C15]-AME,U-[13C22]-TEN,U-[13C10]-TEA)/甲醇 (AOH 6ppm、AME 0.6ppm、TEN 3µg/mL,TEA 15 µg/mL

 · 2 分钟阅读

1. General information

This document is designed and the certified value(s) and uncertainty(ies) are determined in accordance with ISO Guide 31[1].

2.1 Intended use of the RM

- for laboratory use only

- calibration of analytical instruments。

2.2 Instruction for the correct use of the RM  

The compound should be stored at 2-8℃ in a dark place. Before usage of the RM , the compound should be allowed to warm to temperature(20±3℃). The recommended minimum sub - sample amount for all kinds of application is 100 μL. The expiry date of this RM is based on the current knowledge and holds only for proper storage conditions in the originally closed flasks/ packages.

2.3 Hazardous situation

The normal laboratory safety precautions should be observed when working with this RM.Further details for the handing of this RM are available as safety data sheet.

Hazardous Ingredients     Concentration in%     Pictograms     Signal word     Hazard statement(s)

Acetonitrile                         >99.9                                               Danger         H225,H302,H312,H319,H332

3. Certified values and their uncertainties

U-[13C14]-Alternariol,U-[13C15]-Alternariol monomethyl ether,U-[13C22]-Tentoxin,U-[13C10]-Tenuazonic acid in Methanol
CompoundMass concentration a
Certified value bUncertainty c
U-[13C14]-Alternariol6.05μg/mL±0.06μg/mL
U-[13C15]-Alternariol monomethyl ether0.604μg/mL±0.01μg/mL
U-[13C22]-Tentoxin3.00μg/mL±0.06μg/mL
U-[13C10]-Tenuazonic acid15.00μg/mL±0.27μg/mL
a   Values are based on preparation data and confirmed experimentally by HPLC-DADb   Mass concentration based on weighed amount ,purity and dilution step
c   Expanded uncertainty U(k=2) of the value uc according to GUM[3]


4.Discussion of traceability

This calibrant is certified on the basis of gravimetric preparation[3]. Thus the certified value(mass concentration of U-[13C14]-Alternariol,U-[13C15]-Alternariol monomethyl ether,U-[13C22]-Tentoxin,U-[13C10]-Tenuazonic acid is based on the weighed amount of the starting material and is therefore traceable to the stated purity of the solid raw material. High purity material represents a practical realization of concentration units, through conversion of mass to molar quantity.

5. Confirmation of certified value by HPLC-DAD

The certified concentration of the gravimetric prepared solution was confirmed by HPLC-DAD against an independently prepared reference batch of unlabeled Alternariol,Alternariol monomethyl ether,Tentoxin,Tenuazonic acid.

6.Further information

The purchaser must determine the suitability of this product for its particular use. Pribolab makes no warranty of any kind, express or implied, other than its products meet all quality control standards set by Pribolab. We do not guarantee that the product can be used for a special application.  

Inspected by

Quality System Specialist




No comments yet.

添加评论
Ctrl+Enter添加评论