Pribolab - STD#1081-5Pribolab®2µg/mL黄曲霉素(Aflatoxin)B1,G1,0.5µg/mL黄曲霉素B2,G2/乙腈-68619

Pribolab®2µg/mL黄曲霉素(Aflatoxin)B1,G1,0.5µg/mL黄曲霉素B2,G2/乙腈

 · 4 分钟阅读

Certificate of Analysis — STD#1082-5.0ppm

1. General Information

This document is designed and the certified value(s) and uncertainty(ies) are determined in accordance with ISO Guide 31.^[1]^

2.1 Intended Use of the RM

  • For laboratory use only
  • Calibration of analytical instruments

2.2 Instruction for Correct Use

The compound should be stored at -20℃ or below in a dark place. Before usage of the RM,the compound should be allowed to warm to temperature(20±3℃). The recommended minimum sub - sample amount for all kinds of application is 100 μL. The expiry date of this RM is based on the current knowledge and holds only for proper storage conditions in the originally closed flasks/ packages.

2.3 Hazardous Situation

IngredientConcentrationSignal wordHazard statements
Acetonitrile >99.9% ⚠ Danger H225, H302, H312, H319, H332

3. Certified values and their uncertainties

Aflatoxin Total in Acetonitrile
Compound Mass concentration a Compound Mass concentration a
Certified valueb Uncertaintyc Certified valueb Uncertaintyc
Aflatoxin B1 2.00μg/mL ±0.02μg/mL Aflatoxin G1 2.00μg/mL ±0.04μg/mL
Aflatoxin B2 0.500μg/mL ±0.005μg/mL Aflatoxin G2 0.500μg/mL ±0.01μg/mL
a Mass concentration based on weighed amount, purity and dilution steps
b Values are based on preparation data and confirmed experimentally by HPLC-DAD
c Expanded uncertainty U(k=2) of the value uc according to GUM[2]

3. Calculation of uncertainty

After the concentration of the gravimetric prepared solution was confirmed by HPLC-DAD,the uncertainty of the calibrant was calculated on the basis of preparation[3].

Uncertainty components Description Standard uncertainty/uc
Purity(P) of solid
Aflatoxin B1
Aflatoxin B2
Aflatoxin G1
Aflatoxin G2
P1=99.5 ± 0.5%
P2=99.6 ± 0.4%
P3=98.7 ± 1.3%
P4=98.5 ± 1.5%
u(P1)=0.3%↓
u(P2)=0.2%
u(P3)=0.8%
u(P4)=0.9%
Weighing procedure: Weighted sample
maflatoxin,B1=5.025mg
maflatoxin,B2=1.255mg
maflatoxin,G1=5.066mg
maflatoxin,G2=1.269mg
U(m)=0.0000008g+1.30*10-5*mToxin
u(m)=U(m)/2
u1m=0.0004mg
Dilution procedure
Volumetric flask1:Vf1=250mL
Volumetric flask2:Vf2=250mL
One-mark glass pipette:Vp=25mL
calibration flask1: 250mL±0.15mL
repeatability flask1: 0.03mL
volume expansion solvent flask1

calibration flask2: 250mL±0.15mL
repeatability flask2: 0.03mL
volume expansion solvent flask2

Calibration pipette: 25mL±0.03mL
volume expansion solvent pipette
u(cal1)=0.06mL
u(rep1)=0.03mL
u(Vol.exp.1)=0.59mL
u(V1)=0.59mL

u(cal2)=0.06mL
u(rep2)=0.03mL
u(Vol.exp.2)=0.59mL
u(V2)=0.59mL

u(cal3)=0.01mL
u(Vol.exp.3)=0.06mL
u(V3)=0.06mL

a Maximum tolerance of purity was divided by √3

b Calculation of this u-value is based upon the uncertainty formula for the weighed amount as given in the calibration report from annual balance calibration

c,d,e A triangular distribution(division by √6 ) was chosen for the calculation of u(cal)a

d,h Based on a series of ten fill and weigh experiments on a typical 250mL flask; the value was used directly as a standard deviation

e,j,l Based on the density of 0.7857 g/cm3 at temperature T=20℃ and a maximum temperature variation of ±3℃,of volume expansion, relative volume expansion coefficient of acetonitrile is 1370*10-6/℃[4], volume expansion term(rectangular distribution) was divided by √3

f,i,m The three contributions are combined to give the u(V)=√[u(cal)2+u(rep)2+u(Vol.exp)2]

Calculation of the combined uncertainty uc and the expanded standard uncertainty U

CToxin = 10 × mws × P1 × Vp / Vf1 × Vf2 = 10 × 5.025 × 99.5 × 25 / 250 × 250 = 2.00 mg / L

uc(CToxin) = √[ (u(P1)/P1)2 + (u(m)/mws)2 + (u(Vf1)/Vf1)2 + (u(Vf2)/Vf2)2 + (u(Vp)/Vp)2 ] = √[ (0.3/99.5)2 + (0.0004/5.025)2 + (0.59/250)2 + (0.59/250)2 + (0.06/25)2 ] = 0.005

uc(CToxin) = CToxin × 0.005 = 2.00 × 0.005 = 0.01 mg / L

calculation of expanded standard uncertainty U using a coverage factor k=2

U(CToxin) = uc(CToxin) × 2 = 0.01 × 2 = 0.02 mg / L = 0.02 μg / mL

4. Discussion of Traceability

This calibrant is certified on the basis of gravimetric preparation[3]. Thus the certified value(mass concentration of Aflatoxin Total is based on the weighed amount of the starting material and is therefore traceable to the stated purity of the solid raw material. High purity material represents a practical realization of concentration units, through conversion of mass to molar quantity.

5. Confirmation by HPLC-DAD

The certified concentration of Aflatoxin Total of the gravimetric prepared solution was confirmed by HPLC-DAD against an independently prepared reference batch of Aflatoxin Total.

6. Further Information

The purchaser must determine the suitability of this product for its particular use. Pribolab makes no warranty of any kind, express or implied, other than its products meet all quality control standards set by Pribolab. We do not guarantee that the product can be used for a special application.

Inspected by